πŸ”§ Samsung XH40 Image Interference β€” Troubleshooting CaseA Samsung XH40 was experiencing image interference when using th...
12/10/2026

πŸ”§ Samsung XH40 Image Interference β€” Troubleshooting Case
A Samsung XH40 was experiencing image interference when using the abdominal probe.
After inspection, the probe itself was ruled out as the cause. Our technicians found that the interference was related to the working environment. After proper grounding and eliminating the interference source, the ultrasound image returned to normal.
βœ… Fault diagnosis
βœ… Interference troubleshooting
βœ… Ultrasound system repair & technical support
🌍 Worldwide service available
Need help with your ultrasound equipment? Contact us for professional technical support.
πŸ“² WhatsApp: +86 19566042918
πŸ“§ Email: [email protected]
🌐 Rongtao Medical: www.rongtaomedical.com

πŸ”§ Philips Epiq 5 – Error 031 Repair CaseThe system powered on normally, but the probes could not be recognized, and Erro...
06/10/2026

πŸ”§ Philips Epiq 5 – Error 031 Repair Case
The system powered on normally, but the probes could not be recognized, and Error 031 appeared when switching probes.
After testing the probes and each interface, our engineers ruled out probe failure. Further hardware diagnostics indicated a CB Board / ACQ module issue.
βœ… Repair Solution:
The ACQ module was replaced, and the system returned to normal operation.
We provide professional ultrasound equipment troubleshooting, parts replacement, and repair services for customers worldwide.
πŸ“² WhatsApp: +86 19566042918
πŸ“§ Email: [email protected]
🌐 Website: www.rongtaomedical.com

πŸ”§ Philips EPIQ A50  Error 005 When Switching ProbesA Philips EPIQ A50 reported Error 005 when switching probes during op...
05/10/2026

πŸ”§ Philips EPIQ A50 Error 005 When Switching Probes
A Philips EPIQ A50 reported Error 005 when switching probes during operation.
πŸ” Fault Analysis
After running the system self-check, a CB error was detected. Further diagnosis indicated a fault in the ACQ module.
βœ… Repair Result
The faulty ACQ module was replaced, and the system returned to normal operation. The probe switching function was successfully restored.
We provide professional ultrasound equipment diagnosis, repair, parts replacement, and technical support for customers worldwide.
πŸ“² Tel WhatsApp +86 19566042918
πŸ“§ Email [email protected]
🌐 Website www.rongtaomedical.com
Worldwide Ultrasound Repair Support 🌍

πŸ”§ Philips Affiniti50 Repair Case | No Response When Power Button Is PressedA Philips Affiniti50 showed no response when ...
04/10/2026

πŸ”§ Philips Affiniti50 Repair Case | No Response When Power Button Is Pressed
A Philips Affiniti50 showed no response when the power button was pressed.
During inspection, our engineer noticed a burning smell and carefully checked the internal boards. A burnt component was found on the power board, identifying the source of the failure.
βœ… Repair Solution: Power board replacement
βœ… Result: Machine starts normally
βœ… Self-Test: Passed
βœ… Fault Eliminated
We provide professional ultrasound equipment repair, parts replacement, and technical support for customers worldwide.
πŸ“² WhatsApp: +86 195 6604 2918
πŸ“§ Email: [email protected]
🌐 www.rongtaomedical.com

πŸ”§ Mindray DC-3 | Image Missing Repair CaseA Mindray DC-3 was experiencing an image missing abnormality during operation....
02/10/2026

πŸ”§ Mindray DC-3 | Image Missing Repair Case
A Mindray DC-3 was experiencing an image missing abnormality during operation.
After testing multiple probes and confirming the same issue, our technicians diagnosed the problem as a Transmitter & Receiver Board failure.
βœ… Fault: Image missing
πŸ” Diagnosis: Transmitter & Receiver Board failure
πŸ› οΈ Solution: Transmitter Board replacement
✨ Result: Ultrasound image restored to normal
Rongtao Medical provides professional ultrasound equipment troubleshooting, board repair, parts replacement, and technical support for customers worldwide.
πŸ“² WhatsApp: +86 19566042918
☎️ Tel: +86 19928498169
πŸ“§ Email: [email protected]
🌐 Website: www.rongtaomedical.com
Have a similar ultrasound fault? Send us the model and fault details for technical support.

Mindray Resona 62D | Gain K**b Repair CaseThe Gain k**b could be pressed normally, but there was no response when rotati...
01/10/2026

Mindray Resona 62D | Gain K**b Repair Case
The Gain k**b could be pressed normally, but there was no response when rotating it, making gain adjustment impossible.
πŸ” Fault Diagnosis:
After inspection and testing, the fault was traced to the k**b encoder circuit board.
βœ… Repair Result:
The faulty k**b encoder circuit board was replaced. After repair and testing, the 2D Gain control returned to normal operation.
Rongtao Medical provides professional ultrasound equipment repair, parts replacement, and technical support for customers worldwide.
πŸ“ž Tel: +86 2031023254
πŸ“² WhatsApp: +86 19566042918
πŸ“§ [email protected]
🌐 www.rongtaomedical.com
Guangzhou Rongtao Medical Tech LTD

πŸ”§ Madison X6 Image Interference – Repair CaseMadison X6 showed image interference on both abdominal and superficial prob...
29/09/2026

πŸ”§ Madison X6 Image Interference – Repair Case
Madison X6 showed image interference on both abdominal and superficial probes.
After troubleshooting, the issue was identified as environmental interference. After disconnecting the external trolley charging plug, the image returned to normal.
βœ… Fault identified
βœ… Problem solved
🌍 Worldwide ultrasound repair support
πŸ“² WhatsApp: +86 19566042918
πŸ“§ [email protected]

πŸ”§ Samsung H60 Ultrasound System Repair CaseThe Samsung H60 could not start up normally. After pressing the power button,...
23/09/2026

πŸ”§ Samsung H60 Ultrasound System Repair Case
The Samsung H60 could not start up normally. After pressing the power button, the system repeatedly powered off and restarted.
βœ… Fault analysis: Mainboard / power supply issue
βœ… Mainboard replaced and troubleshooting completed
βœ… System restored to normal operation
βœ… Tested after repair
Professional ultrasound equipment repair service with worldwide support. 🌍
πŸ“© Contact us for repair inquiries:
WhatsApp: +86 19566042918
Email: [email protected]
www.rongtaomedical.com

πŸ”§ Siemens SC2000 Ultrasound System Repair CaseThe Siemens SC2000 could not start normally and repeatedly lost power and ...
23/09/2026

πŸ”§ Siemens SC2000 Ultrasound System Repair Case
The Siemens SC2000 could not start normally and repeatedly lost power and restarted after power supply replacement.
After inspection, the power indicator was normal. Further troubleshooting revealed multiple bent pins in the motherboard interface socket, affecting the PC connection.
βœ… Power supply replaced
βœ… Motherboard interface pins repaired and aligned
βœ… PC connection restored
βœ… System tested and operating normally
Professional ultrasound equipment repair and technical support with worldwide service. 🌍
πŸ“© Contact us for repair inquiries:
WhatsApp: +86 19566042918
Email: [email protected]
Website: www.rongtaomedical.com

πŸ”§ Toshiba Aplio 500 Ultrasound Repair CaseThe system displayed β€œSystem Busy” after startup and could not operate normall...
23/09/2026

πŸ”§ Toshiba Aplio 500 Ultrasound Repair Case
The system displayed β€œSystem Busy” after startup and could not operate normally.
πŸ” Fault Analysis:
Initial inspection indicated possible issues with the PC and system.
πŸ›  Repair Solution:
βœ… PC replacement
βœ… System replacement
βœ… System testing after repair
After repair, the fault was eliminated and the Toshiba Aplio 500 started and operated normally again.
Professional ultrasound equipment repair service with worldwide support. 🌍
πŸ“© Contact us for repair inquiries:
WhatsApp: +86 19566042918
Email: [email protected]

Address

Room 201, 203, Building C , No. 26, Ningcai Road, Huangpu District
Guangzhou
510530

Opening Hours

Monday 09:00 - 19:00
Tuesday 09:00 - 19:00
Wednesday 09:00 - 19:00
Thursday 09:00 - 19:00
Friday 09:00 - 19:00

Telephone

+8619566042918

Alerts

Be the first to know and let us send you an email when Guangzhou Rongtao Medical TECH LTD posts news and promotions. Your email address will not be used for any other purpose, and you can unsubscribe at any time.

Contact The Practice

Send a message to Guangzhou Rongtao Medical TECH LTD:

Shortcuts

Share