The Standard Operating Procedure (SOP) for operating the Systonic S-935 Microprocessor Flame Photometer involves setup, calibration, sample testing, and shutdown.
1. Pre-Check and Setup
• Connect the Systonic S-935 to a stable power supply of 230 VAC, 50 Hz.
• Attach the clean, dry oil-free air compressor and the LPG gas supply line to the rear inlets.
• Ensure the appropriate filters (Sodium and Potassium are standard; Calcium and Barium are optional) are properly placed.
2. Ignition and Stabilization
• Turn on the air compressor and adjust the air pressure to the recommended operating level.
• Turn on the LPG gas supply and adjust the gas control k**b slightly.
• Power on the instrument using the main switch. Use the auto-ignition system to light the flame.
• Allow the instrument to stabilize with distilled water aspirating through the atomizer for 10 to 15 minutes.
3. Calibration
• Access the calibration menu on the graphical LCD screen.
• Perform a blank adjustment by aspirating distilled water to set the zero reading.
• Calibrate the instrument using up to 5 standard solutions (e.g., standard concentrations of Na and K) via the built-in curve fitting software.
4. Sample Measurement
• Aspirate your prepared liquid test samples one by one through the axial flow atomizer.
• Read and record the stable concentration values (in ppm or meq) directly from the backlit graphical LCD.
• Aspirate distilled water between samples to clean the nebulizer and prevent cross-contamination or clogging.
5. Shutdown Procedure
• Aspirate distilled water for at least 5 minutes after finishing all measurements to clear any salt residues.
• Turn off the LPG gas valve first to extinguish the flame safely.
• Turn off the air compressor and switch off the main power to the unit.
If you need specific details on standard preparation recipes, troubleshooting low readings, or maintenance guidelines, let me know how to proceed!